Philips IntelliVue MP5 Specifications
Dimensions
- Height: 9.76″ (24.8 cm)
- Width: 10.2″ (25.9 cm)
- Depth: 7.32″ (18.6 cm)
- Weight: 8.8 lbs (4 kg)
Power
- Power Consumption: < 40W average; < 65W Peak
- Line Voltage: 100 to 240 V ~
- Current: 1.3 to 0.7A
- Frequency: 50/60 Hz
Display
- Type: SVGA
- Size: 8.4″ (21.3 cm)
- Resolution: 800 x 600
- Refresh Rate: 60 Hz
Battery
- Type: Lithium-Ion
- Capacity: 6000 mAh
- Voltage: 10.8 V
- Weight: 480g (Per battery)
- Charge Time: 4 hours (When off); 5 hours or more when on
- Operating Time: 4 Hours (Basic monitoring configuration); 3 hours (Extended monitoring configuration)
Operating Environmental Specifcaitons
- Temperature: 32 to 104°F (0 to 40°C)
- Humidity: 15 to 95% non condensing
- Altitude: 10,000 ft (-500 to 3000 m)
Storage Environmental Specifcaitons
- Temperature: -4 to 140°F (-20 to 60°C)
- Humidity: 5 to 90% non condensing
- Altitude: 15,000 ft (-500 to 4600 m)
Sweep Speeds
- 6.25, 12.5, and 50 mm/s with +/- 5% accuracy
CO2
- Range: 0 to 150 mmHg (0 to 20.0 kPa)
- Accuracy: after 2 minutes warmup: For values between 0 and 40 mmHg:±2.0 mmHg (±0.29 kPa) For values from 41 to 70 mmHg: ±5% of reading For values from 71 to 100 mmHg: ±8% of reading The specifications are valid for standard gas mixtures, balance air, fully hydrated at 35°C, Pabs = 760 mmHg, flow rate = 2 l/min
- Resolution: Numeric: 1.0 mmHg (0.1 kPa); Wave: 0.1 mmHg (0.01 kPa)
- Stability: Short-term drift: ±0.8 mmHg over four hours; Long-term drift: Accuracy specification will be maintained over a 120 hour period
awRR
- Range: 2 to 150 rpm
- Accuracy: ±1 rpm
Warm-up Time
- 2 minutes with CO2 transducer attached for full accuracy specification
Response Time
- Less than 60 ms
Temperature
Temp
- Range: –1 to 45 °C (30 to 113 °F)
- Resolution: 0.1 °C (0.2 °F)
- Accuracy: ±0.1 °C (±0.2 °F)
Average Time Constant
- Less than 10 seconds
Alarms
- Range: –1 to 45 °C (30 to 113 °F)
- Adjustment: -1 to 35 °C (30 to 95 °F): 0.5 °C (1.0 °F) steps 35 to 45 °C (95 to 113 °F): 0.1 °C (0.2 °F) steps
Invasive Pressure Performance
Measurement Rnage
- –40 to 360 mmHg
Pulse Rate
- Range: 25 to 350 bpm
- Accuracy: ±1% Full Range
- Resolution: 1 bpm
Input Sensitivity
- Sensitivity:5µV/V/mmHg (37.5µV/ V/kPa)
- Adjustment range:±10%
Transducer
- Load Impedance:200 to 2000 Ω (resistive)
- Output Impedance:≤3000 Ω (resistive)
Frequency Response
- dc to 12.5 Hz or 40 Hz
Zero Adjustment
- Range: ±200 mmHg (±26 kPa)
- Accuracy: ±1 mmHg (±0.1 kPa)
- Drift: Less than 0.1mmHg/°C (0.013 kPa/ °C)
NBP Performance Specifications
Measurement Ranges
- Systolic: Adult: 30 to 270 mmHg (4 to 36 kPa); Pedi: 30 to 180 mmHg (4 to 24 kPa); Neo: 30 to 130 mmHg (4 to 17 kPa)
- Diastolic: Adult: 10 to 245 mmHg (1.5 to 32 kPa); Pedi: 10 to 150 mmHg (1.5 to 20 kPa); Neo: 10 to 100 mmHg (1.5 to 13 kPa)
- Mean: Adult: 20 to 255 mmHg (2.5 to 34 kPa); Pedi: 20 to 160 mmHg (2.5 to 21 kPa); Neo: 20 to 120 mmHg (2.5 to 16 kPa)
- Pulse Rage: Adult:40 to 300; Pedi: 40 to 300; Neo: 40 to 300
Accuracy
- Max. Std. Deviation: 8 mmHg (1.1 kPa)
- Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Measurement Accuracy
- 40 to 100 bpm: ± 5 bpm
- 101 to 200 bpm: ± 5% of reading
- 201 to 300 bpm: ± 10% of reading (average over NBP measurement cycle)
Heart Rate Range
- 40 to 300 bpm
Measurement Time
- Typical at HR > 60bpm
- Auto/manual: 30 seconds (adult) 25 seconds (neonatal) Stat: 20 seconds
- Maximum time: 180 seconds (adult/pediatric) 90 seconds (neonates)
Cuff Inflation Time
- Typical for normal adult cuff: Less than 10 seconds
- Typical for neonatal cuff: Less than 2 seconds
SpO2 Performance Specifications
Pulse
- Range: 30 to 300 bpm
- Accuracy: ±2% or 1 bpm, whichever is greater
- Resolution: 1 bpm
Sensors
- Wavelength range: 500 to 1000 nm
- Emitted Light Energy: ≤ 15mW Information about the wavelength range can be especially useful to clinicians (for instance, when photodynamic therapy is performed)
Pulse Oximeter Calibration Range
- 70 – 100%
ECG Arrhythmia and ST Performance
Cardiotach
- Range: Adult/pedi: 15 to 300 bpm; Neo range: 15 to 350 bpm
- Accuracy: ±1% of range
- Resolution: 1 bpm
- Sensitivity: ≥200 µVpeak
PVC Rate
- Range: 0 to 300 bpm
- Resolution: 1 bpm
ST Numeric
- Range: -20 to +20 mm
- Accuracy: ±0.5 mm or 15%, whichever is greater
- Resolution: 0.1 mm
Sinus and SV Rhythm Ranges
- Brady: Adult: 15 to 60 bpm; Pedi: 15 to 80 bpm; Neo: 15 to 90 bpm
- Normal: Adult:60 to 100 bpm; Pedi: 80 to 160 bpm; Neo: 90 to 180 bpm
- Tachy: Adult: > 100 bpm; Pedi: >160 bpm; Neo: >180 bpm
Bandwidth
- Diagnostic Mode: Adult/neo/pedi: 0.05 to 150Hz
- Extended Monitoring Mode: Neo/pedi: 0.5 to 150Hz
- Monitoring Mode: Adult: 0.5 to 40Hz; Neo/pedi: 0.5 to 55Hz
- Filter mode: Adult/neo/pedi: 0.5 to 20H
HR
- Range: 15 to 300 bpm maximum delay: 10 seconds according to AAMI EC 13- 1992 standard