Specifications
HeartStart Dimensions
- Height: 7.5 in (19 cm)
- Width: 14.8 in (37.6 cm)
- Depth: 13.7 in (34.6 cm)
- Weight: 13.2 lbs (6 kg)
HeartStart Battery
- Type: HeartStart Rechargeable Sealed lead acid
- Operating Time: About 100 minutes
- Charging Time: About 14.5 hours
- Power: 2 Ah, 1
HeartStart Display (EL)
- Type: EL – Electroluminescent
- Size: 4.5″ x 3.4″ (11.5 cm x 8.6 cm)
- Resolution: 320 x 240
- Sweep Speed: 29 mm/s nominal
- Viewing Time: 4 Seconds
HeartStart Display (LCD)
- Type: LCD – TFT Color Liquid Crystal Display
- Size: 4.3″ x 3.3″ (11.1 cm x 8.4 cm) on the HeartStart
- Resolution: 320 x 240
- Sweep Speed: 29 mm/s nominal
- Viewing Time: 4 Seconds
Power
- AC Line Powered: 100-240 VAC, 50/60 Hz, 1.5 A
HeartStart ECG Monitoring
- Size: 2.5, 5, 10, 20, 40 mm/mV
- Heart Rate: 15 to 300 bpm
- Inputs: 3 or 5-lead patient cable
SpO2 Monitoring
- Range: SpO2: 0 to 100%; Pulse Rate: 30 to 300 bpm
- Pulse Rate Accuracy: 2% or 1 bpm (whichever is greater)
- Wavelength Range: 500 to 1000 nm
- Emitted Light Energy: ≤ to 15 mW
- Dispaly Update Interval: ≤ 60 seconds
- Resolution: 1%
Synchronized Cardioversion
- Yes in manual mode
Noninvasive Pacing on the
- Waveform: Monophasic Truncated Exponential
- Current Pulse Amplitude: 10 mA to 200 mA (5 mA Resolution); Accuracy 10 mA to 50mA ± 5mA, 50 mA – 200mA ± 10%
- Pulse Width: 20 ms with accuracy of +0, -5 ms
- Rate: 30 ppm to 180 ppm (10ppm increments); accuracy ± 1.5%
- Modes: Demand or Fixed Rate
- Refractory Period: 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm)
Event Storage
- Internal storage up to 300 events and 50 ECG strips at 11 seconds each
- Events can be marked with a Mark symbol and, if configured for dug annotation, the following labels can be added: Epinephrine (Adrenaline in U.K. & Australia, Atropine, Lidocaine, or other.
- Data Card Event Summary: One Data Card can store a minimum of 2 hours of continuous ECG waveforms and events.
Thermal Array Printer
- Continuous Real-Time Strip: User starts and stops the strip. The strip prints the selected ECG lead.
- Event Printing: Mark key automatically documents ECG and events during defibrillation episodes. The Mark key can annotate the event with one of the following labels: Epinephrine (Adrenaline), Atropine, Lidocaine, and other
- Auto Printing: The printer can be configured to automatically print on Mark, Charge, Shock, and Alarm
- Delayed Printing: The printer can be configured to run real-time or with a 6-second delay
- Speed: 25 mm/s with an accuracy of ± 5%
- Amplitude Accuracy: ± 10% or ± 50 uV, whichever is greater
- Paper Size: 50 mm by 30 m (100 ft)