GE Versamed iVent 201 Specifications
Dimensions
- Height: 13″ (33 cm)
- Width: 9.5″ (24 cm)
- Depth: 10.3″ (26 cm)
- Weight: Up to 27.8 lb (12.6 kg)
Ventilation Modes
- Assist/Control (A/C)
- Volume Controlled (A/C)
- Pressure Controlled (A/C)
- Synchronized Intermittent Mandatory Ventilation (SIMV)
- Volume Controlled SIMV (Vctrl)
- Pressure Controlled SIMV (Pctrl)
- Continuous Positive Airway Pressure (CPAP) with Pressure Support Ventilation (PSV)
O2 Supply Specifications
- High Pressure: Pressure Range 40 –75psi (2.8-5.1 bar). Warning: The exact oxygen pressure which the iVent 201 uses depends on the version you have. Check the back label for the pressure levels of your unit.
- Low Pressure: Maximum flow 15L/min or 0.5 psi
Ventilation Performance and Parameters
- Respiratory Rate: 1 – 12 ±1, 12 – 80 ± 2 bpm
- Tidal Volume: 50-2000 mL
- Accuracy of Tidal Volume Delivery: ±10% or ±10 mL whichever is greater
- Accuracy of Exhale Tidal Volume Measurement: ±15% above 100 mL from actual reading or ±10 cc below 100 mL
- Inspiratory Pressure Limit: 5 to 80 ±5 cmH2O
- Inspiratory Time: Adaptive Time or 0.2 to 3 ±10% sec
- Peak Flow (PIF): Adaptive Flow or to 120 ±10% L/min Spontaneous to 180 ±10% L/min
- Oxygen Mix (FiO2 ): 21% to 100% ±10% of the set value or ±3% absolute, whichever is greater
- PEEP: 0 to 40 ±1 cmH2O or ±10%, whichever is greater
- Trigger Sensitivity: 1 to 20 L/min Flow, Off – 0.5 to – 20 cmH2O Pressure, Off
- PSV: 0 to 60 ±10% cmH2O
- Positive Pressure Relief Valve: 80 cmH2O
- Pneumatic Nebulizer: 5 to 240 minutes
- Controlled Pressure: 5 to 80 cmH2O ±5 cmH2O
- FiO2 at Power up: 21%, 40%, 60%, 100% (configurable)
- Configurable Humidifier Setting (Purging Cycle): HME (10 minutes), Heated (1 minute), Off
Power Supply
- External AC: 100 to 240 V, 50 to 60 Hz, Max 1.6 A
- External DC: 12 to 15 V, Max 8.5 A
- Internal battery: Sealed lead-acid 12 V (7.8 Ah) (rechargeable)
- Operating time: Depending on ventilator settings and impedance – standard internal up to 2 hours, optional extended internal up to 4 hours, and optional external up to 8 hours.
Oxygen Supply
- High-pressure supply: 40 to 75 psi (2.8 to 5.1 bar)
- Low pressure: Max 15 L/min or 0.5 psi
Environmental
- Operating temperature: 0 to 50° C/32 to 120° F
- Storage temperature: –15 to 70° C/–4 to 140°F
- Relative humidity: 15 to 95% at 30° C/85° F
- Water/dust resistance: IP54 (splash proof)
- Atmospheric pressure: 430 to 825 mm Hg (15,000 feet)
- Vibration:
- IEC 68-2-6 and IEC 68-2-34
- mIL-STD-810E
- Shock: Total external sound level: 40 to 45 dBa at one meter
- IEC 68-2-27 (100 g)
- MIL-STD-810E
Ventilation Performance and Controlled Parameters
- Respiratory rate: 1 to 80 bpm
- Tidal volume: 50 to 2,000 ml
- Inspiratory pressure limit: 5 to 80 cm H2O
- Inspiratory time: Adaptive I-Time* or 0.2 to 3 sec
- Esens: 10-90%
- Rise time: Mid/high/max/auto
- Peak flow: Adaptive Flow* or 1 to 120 L/min; Spontaneous up to 180 L/min
- Oxygen mix (FiO2): 21 to 100%
- PEEP: 0 to 40 cm H2O
- Trigger sensitivity: 1 to 20 L/min flow sensitivity –0.5, to –20 cm H2O pressure sensitivity
- PSV: 0 to 60 cm H2O
- Positive pressure relief valve: 80 cm H2O
- Synchronized nebulizer: 5 to 240 min
Monitoring and Displayed Parameters
- Airway pressure (analog bar graph & numerical)
- Total breath rate
- I:E ratio
- Exhaled tidal volume
- Exhaled minute volume
- Peak flow
- Inspiratory time
- Electrical power source (AC/external DC/internal DC)
- Battery level
- Pressure, flow, and volume Waveforms software package: Trending of monitored data (72 hours)
- Real-time pressure and flow waveforms
- Waveform history brows
- Respiratory mechanics (C, R, MAP, RR/Vt)SpO2
- Pressure, flow and volume loops
Customer reviews
Reviews
There are no reviews yet.
Write a customer review