Drager Savina Specifications
Dimensions (without Trolley)
- Height: 156.1 in (383 mm)
- Width: 15 in (180 mm)
- Depth: 14.1 in (358 mm)
- Weight (basic device): 53 lbs (24 kg)
- Diagonal Screen Size: 6.1 in TFT color Screen
Measured value display
- Airway pressure measurements: Peak pressure, plateau pressure, mean airway pressure, PEEP 0 -100 mbar (cmH2O)
- Minute Volume (MV): Total MV, spontaneous MV 0 to 99 L/min, BTPS
- Tidal Volume VT: Inspiratory VT, expiratory VT 0 to 3999 mL, BTPS
- Breathing Frequency: Total and spontaneous breathing frequency, 0 -150 bpm
- Inspiratory O2-concentration: 21 to100 Vol. %
- Breathing gas temperature: 18 to 48 °C (sensor optional)
- Curve displays: Airway pressure / time, flow / time
- Ventilation Ratio (I:E):
Performance Data
- Maximum flow for pressure assist/spontaneous breathing: 180 L/min
- Value response to TO..90: ≤5 ms
- Control principle: time-cycled, volume-constant, pressure-controlled
- Safety valve opening pressure: 100 mbar (cmh2O)
- Emergency value: automatically enables spontaneous breathing with filtered ambientair if air and O2 supply should fail.
- Automatic gas switch-over function if O2 supply fails
- Output for pneumatic medicament nebuliser : Synchronized with inspiration
Operating Data
- Main Power connection: 100 V to 240 V, 50/60 Hz AC, 10 to 36 V DC
- Typical Power Consumption: 100W
- Internal battery: Approx. 60 min ( optional extension up to 7h)
Gas Supply
- Air: Turbine technology
- O2 Gas Supply:
Ventilation Modes
- IPPV (CMV), IPPVAssist (CMVAssist)
- SIMV, SIMVASB (SIMV/PS)
- CPAP, CPAPASB (CPAP/PS)
- BIPAP1), (PCV+) (optional), BIPAP)ASB (PCV+/PS) (optional)
Enhancements
- NIV – Non Invasive Ventilation with optimized alarm system and automatic leakage compensation (optional)
- AutoFlow®
- Automatic adaptation of the inspiratory flow in volume orientated ventilation modes (optional)
- LPO – Low Pressure Oxygen. Independant oxygen supply, e.g. with an O2 concentrator (optional)
- Graphic screen – Advanced ventilation monitoring (optional)
- Nurse call – Connection for transmitting alarm signals to a central alarm system (optional)
Customer reviews
Reviews
There are no reviews yet.