GE Avance CS2 Specifications
Physical Specifications
- Height: 54″ (139 cm)
- Width: 30″ (77 cm)
- Depth: 29.9″ (76 cm)
- Weight: 324.08 lbs (147 kg)
Top Shelf
- Weight Limit: 75 lbs (34 kg)
- Width: 27.4″ (69.7 cm)
- Depth: 17.3″ (44 cm)
Work Surface
- Height: 32.2 in (81.7 cm)
- Size: 409 in2 (2640 cm2)
Casters
- Diameter: 5.2″ (13 cm)
- Brakes: Central Brake
Modes of Ventilation – Standard
- VCV (Volume Control) Mode with tidal volume compensation
Modes of Ventilation – Optional
- Pressure Control
- Pressure Controlled Ventilation-Volume Guaranteed (PCV-VG)
- SIMV (Synchronized Intermittent Mandatory Ventilation) (volume and pressure)
- PSVPro* (Pressure Support with Apnea backup)
- CPAP+PSV (Pressure support mode)
- SIMV PCV-VG
Ventilator Parameter Ranges
- Tidal volume range: 5 to 1500ml ( Volume Control, PCV-VG and SIMV volume 20 to 1500ml) (PCV modes 5 to 1500ml)
- Incremental settings:
- 20 to 50 mL (increments of 1 mL)
- 50 to 100 mL (increments of 5 mL)
- 100 to 300 mL (increments of 10 mL)
- 300 to 1000 mL (increments of 25 mL)
- 1000 to 1500 mL (increments of 50 mL)
- Minute volume range: Less than 0.1 to 99.9 L/min
- Pressure (Pinspired) range: 5 to 60 cmH2O (increments of 1 cmH2O) 5 to 1500 mL volume delivery
- Pressure (Pmax) range: 12 to 100 cmH2O (increments of 1 cmH2O)
- Pressure (Psupport) range: Off, 2 to 40 cmH2O (increments of 1 cmH2O)
- Rate: 4 to 100 breaths per minute for Volume Control and Pressure Control; 2 to 60 breaths per minute for SIMV, PSVPro and SIMV PCV-VG; 4 to 60 bpm for CPAP+PSV (increments of 1 breath per minute)
- Inspiratory/expiratory ratio: 2:1 to 1:8 (increments of 0.5)
- Inspiratory time: 0.2 to 5.0 seconds (increments of 0.1 seconds) (SIMV, PSVPro and CPAP PSV)
- Trigger window: 0 to 80% (increments of 5%)
- Flow trigger: 1 to 10 L/min (increments of 0.5 L/min) 0.2 to 1 L/min (increments of 0.2 L/min)
- Inspiration termination level: 5 to 75% (increments of 5%)
- Inspiratory Pause range: 0-60%
Positive End Expiratory Pressure (PEEP)
- Type: Integrated, electronically controlled
- Range: OFF, 4 to 30 cmH2O (increments of 1 cmH2O)
Ventilator Performance
- Pressure range at inlet: 240 kPa to 700 kPa (35 psig to 100 psig)
- Peak gas flow: 120 L/min + fresh gas flow
- Flow valve range: 1 to 120 L/min
- Flow compensation range: 150 mL/min to 15 L/min
Delivery Monitoring Accuracy
- Volume delivery:
- > 210 mL = better than 7%
- ≤ 210 mL = better than 15 mL
- < 60 mL = better than 10 mL
- Pressure delivery: ±10% or ±3 cmH2O
- PEEP delivery: ±1.5 cmH2O
- Volume monitoring:
- > 210 mL = better than 9%
- ≤ 210 mL = better than 18 mL
- < 60 mL = better than 10 mL
- Pressure monitoring: ±5% or ±2 cmH2O
Ventilator Components
- Type: Variable orifice flow sensor
- Dimensions: 22 mm OD and 15 mm ID
- Location: Inspiratory outlet and expiratory inlet
Oxygen Sensor
- Type: Optional galvanic fuel cell or paramagnetic with Airway Module option
Ventilator Display
- Display size: 15 inch
- Pixel format: 1024 x 768
Battery Backup
- Backup power: Demonstrated battery time is up to 90 minutes when fully charged.
- Battery time under extreme conditions is 30 minutes.
- Battery type: Internal rechargeable sealed lead acid
Communication Ports
- RS-232C compatible serial interface
- Ethernet
- Datex-Ohmeda device interface solutions port
- USB port
- VGA Output
Anesthetic Agent Delivery
- Anesthesia Vaporizers: Tec* 6 Plus, Tec 7
- Number of positions: 2 (3 as an option)
- Mounting: Tool-free installation Selectatec* manifold interlocks and isolates vaporizers